Column Oven
• Dimensions: 28 × 31 × 16 cm.
Accommodates up to two 105 m ×
0.530 mm id capillary columns or two
10-ft glass packed columns (9 in. coil
diameter, 1/4 in. od), or two 20-ft
stainless steel packed columns
(1/8 in. od).
• Operating temperature range suitable
for all columns and chromatographic
separations. Ambient
temperature +4 °C to 450 °C.
-With LN2 cryogenic cooling:
–80 to 450 °C.
-With CO2 cryogenic cooling:
–40 to 450 °C.
• Temperature setpoint resolution:
0.1 °C.
• Supports 20 oven ramps with
21 plateaus. Negative ramps are
allowed.
• Maximum achievable temperature
ramp rate: 120 °C/min (120 V units
are limited to 75 °C/min, see
Table 1).
• Maximum run time: 999.99 min
(16.7 h).
• Oven cool down (22 °C ambient) 450
to 50 °C in 4.0 min (3.5 min with
oven insert accessory).
• Ambient rejection: < 0.01 °C per 1 °C.
Electronic Pneumatics
Control (EPC)
• Compensation for barometric pressure
and ambient temperature
changes is standard.
• Pressure has typical control of
±0.001 psi for the range of 0 to
150 psi. Pressure setpoints may be
adjusted in increments of 0.001 for
the range 0.000 to 99.999 psi;
0.01 psi for the range 100.00 to
150.00 psi.
• User may select pressure units as
psi, kPa, or bar.
• Pressure/flow ramps: Three
maximum.
• Carrier and makeup gas settings
selectable for He, H2, N2, and
argon/methane.
• Flow or pressure setpoints for each
inlet or detector parameter with both
GC/MS
Ionization mode (standard) EI
Ionization modes (optional) PCI, NCI, EI acquisition with CI source
EI ion source type Four supported sources: Stainless Steel, Inert, Extractor, or High-Effi ciency Source (HES)
Electron energy 5–241.5 eV
Emission current 0–315 μA
CI gases Dual gas inlet
Filaments Dual for EI, single for CI
Transfer line temperature 100–350 °C
Ion source temperature 150–350 °C
Quadrupole temperature 106–200 °C
Mass fi lter Heated monolithic hyperbolic quadrupole
Ion source - mass fi lter interface Dynamically rampable entrance lens
Mass range 1.6 to 1050 u
Mass resolution Unit mass
Mass accuracy 1-μL injection of a 100 pg/μL OFNa standard scanning from 50–300 u will give its monoisotope at m/z 271.987 ± 0.005a
Spectral accuracy 1-μL injection of a 100 pg/μL OFN standard scanning from 50–300 u will give 99.0 % spectral accuracya
Mass axis stability Better than 0.10 u/48 hrs
Detector Series II Triple-Axis Detector with high energy dynode and long life electron multiplier (EM)
Dynamic range (electronic) 106
Scan rateb Agilent 5977B Inert Plus (Extractor source) and Agilent 5977B HES up to 20,000 u/sec,
Agilent 5977B (SS source) up to 12,500 u/sec
SIM 60 ions × 100 groups
Pumping system (helium) 65 L/s Diffusion pump and 255 L/s turbo pump with 2.5 m3/hr mechanical pump
Maximum recommended analytical
gas fl ow
1.5 mL/min (diffusion pump)
4 mL/min (turbo pump)
Recommended maximum sustained
column fl ow to MSc
3 mL/min (diffusion pump)
50 mL/min (turbo pump)
Instrument control Agilent MassHunter data system
Maintenance accessibility Source (including fi laments, lenses), mass fi lterd, and detector on removable plate for easy access
Maintenance scheduling Early Maintenance Feedback (EMF) for GC and MSD, user-defi ned maintenance schedule, with display of current status